Home Monofocal IOLs HOYA Nanex multiSert+ NC1-SP
HOYA Nanex Len front HOYA Nanex Injector with lens HOYA Nanex Len side

Hoya Nanex multiSert+ NC1-SP IOL

PERFORM MICRO INCISION CATARACT SURGERY WITHOUT COMPROMISES

The preloaded Nanex multiSert+system eliminates the compromises traditionally associated with sub-2.2 mm cataract surgery.

Proven hydrophobic acrylic IOL material

Nanex multiSert+ NC1-SP is a clear IOL designed for micro incision cataract surgery as low as 1.8 mm.

Based on HOYA’s proprietary proven hydrophobic acrylic IOL material. To date, over 10 million units have been implanted worldwide, over 15 years.

 

Wound protection

With the world’s smallest nozzle size for an open-loop preloaded hydrophobic IOL system, the Nanex multiSert+ provides the smallest final incision size and minimises wound stretch.

Smooth lens delivery

Nanex multiSert+ is designed to provide a smooth lens delivery. Predictable lens release through HOYA’s unique patented preloaded IOL delivery mechanism ensures outstanding delivery consistency and smoothness.

Sharp optic edge

A sharp optic edge is designed to minimise PCO (Posterior Capsular Opacification).

Significant PCO reduction is achieved through HOYA’s proprietary active oxygen processing treatment to the posterior surface¹.

4-in-1 preloaded IOL system

Providing choice for surgeon preference, the preloaded IOL system enables push or screw injection, and insert shield in advanced or default position.

Nanex multiSert+ application

Download
Technical specifications
Description & Technical Specs
HOYA Nanex multiSert+ NC1-SP
Model Name NC1-SP
Optic design Aspheric design with sharp optic edge
Optic & haptic material Hydrophobic acrylic with UV and blue light filter
Haptic design Modified C-loop, 5° angulation
Diameter (optic/OAL) 6.0 mm/ 13.0 mm
Power +6.00 to +30.00 D (in 0.50 D increments)
Nominal A-Constant* 119.2
Injector multiSert+ preloaded
Front injector tip outer diameter 1.62 mm
Incision size As low as 1.8 mm
Optimised constants† Haigis a0 = -0.2676

a1 = 0.2382 a2 = 0.1993

Hoffer Q pACD = 5.715

Holladay 1 SF = 1.904

SRK/T A = 119.112

¹Matsushima H, et al. Active oxygen processing for acrylic intraocular lenses to prevent posterior capsule opacification. J Cataract Refract Surg. 2006; 32:1035-1040.

* The A-Constant is presented as a starting point for the lens power calculation. When calculating the exact lens power, it is recommended that calculations be performed individually, based on the equipment used and operating surgeon’s own experience.

† These optimised constants for the calculation of intraocular lens power published by IOLCon on their website: https://iolcon.org are calculated from 211 clinical results for Nanex™ multiSert+™ model NY1-SP/NC1-SP as of September 9, 2021. These constants are based on actual surgical data and are provided by IOLCon as a starting point for individual constant optimisations. The information available on the website is based on data originating from other users and not by HOYA Surgical Optics (“HSO”). HSO therefore does not warrant the correctness, completeness and currentness of the contents on the said website.